Analytical performance validation of a coronary heart disease risk assessment multi-analyte proteomic test

Cytokines;Chemokines;细胞因子;趋化因子;MSD;Cytokines;Chemokines
浏览次数:9 分享:

Nolan N, Tee L, Vijayakumar S, Burazor I, Hytopoulos E, Biggs WH, Beggs M, French C, Harrington DS.

  • Expert Opin Med Diagn
  • 2012
  • 7(2):127-36.
  • Human,Mouse,Non-Human Primate
  • MSD
  • Serum
  • 免疫/内分泌
  • CTACK, HGF, IL-16, MCP-3

相关货号

LXMH10-2LXMH10-4LXMH111-1LXMH11-2LXMH13-1LXMH30-1LXMH37-1LXMH40-1LXMH44-1LXMH46-1LXMH54-1LXMH71-1LXMH87-1LXMM10-4LXMM50-1LXMM58-1LXMN09-1LXMN10-3LXMN24-2LXMN61-1

Abstract

Background: Coronary heart disease (CHD) remains prevalent despite efforts to improve CHD risk assessment. The authors developed a multi-analyte immunoassay-based CHD risk assessment (CHDRA) algorithm, clinically validated in a multicenter study, to improve CHDRA in intermediate risk individuals.

Objective: Clinical laboratory validation of the CHDRA biomarker assays' analytical performance.

Methods: Multiplexed immunoassay panels developed for the seven CHDRA assays were evaluated with donor sera in a clinical laboratory. Specificity, sensitivity, interfering substances and reproducibility of the CHDRA assays, along with the effects of pre-analytical specimen processing, were evaluated.

Results: Analytical measurements of the CHDRA panel proteins (CTACK, Eotaxin, Fas Ligand, HGF, IL-16, MCP-3 and sFas) exhibited acceptable accuracy (80 - 120%), cross-reactivity (< 1%), interference (< 30% at high concentrations of bilirubin, lipids, hemoglobin and HAMA), sensitivity and reproducibility (< 20% CV across multiple runs, operators and instruments). Recoveries from donor sera subjected to typical clinical laboratory pre-analytical conditions were within 80 - 120%. The pre-analytical variables did not substantively impact the CHDRA scores.

Conclusions: The CHDRA panel analytical validation in a clinical laboratory meets or exceeds the specifications established during the clinical utility studies. Risk score reproducibility across multiple test scenarios suggests the assays are not susceptible to clinical laboratory pre-analytical and analytical variation.